FDAFDA · Alimentos y medicamentos·N.º oficial Z-1275-2016
Guardian II Hemostasis Valve, models 8210 and 8211. The following models are only available Outside the US: FH101, FH101-T, FH101-25, and FH101-50. Product Usage: The Guardian hemostasis valves are designed to be used as conduits when interventional devices with diameters up to 8 F are inserted into the human vascular system. The devices have two seals: the low-pressure seal, and the high-pressure seal. Closure of the high-pressure seal, which is achieved when the nut can no longer rotate, secures the diagnostic/interventional device in position within the vasculature and allows for pressure injections up to 600 psi.
Publicado el 13 de abril de 2016 · Inicio del retiro: 4 de marzo de 2016
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Vascular Solutions became aware of a potential problem with the click version of the Guardian II hemostasis valves. The low pressure seal may not close properly, which may allow air to be introduced into the device and may lead to risk of an air embolism.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Vascular Solutions, Inc.
- Distribuidor
- Nationwide Distribution in the states of AL, AR, CA, CO, DE, FL, GA, ID, IL IN, IA,KY, LA, MD, MA, MI, MO, NE,NV, NH, NY, NC, OH, PA, SD, TN, TX, UT, VA, WA, and WV. OUS: Ireland
- Lote
- Lot numbers: 41776, 41777, 41874, 42108, 42525, 581766, 581767, 583285, and 583286.
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States