FDAFDA · Alimentos y medicamentos·N.º oficial Z-1370-2016

Boston Scientific, PT Graphix Magnet J-TIP Guidewire with ICE Hydrophilic Coating, Sterile EO, UPN H7491490201MJ2 (Carton Label) ,H7491490201MJo (UPN referenced on the Pouch Label) . Boston Scientific PT Graphix Magnet Guidewire with ICE Hydrophilic Coating is a steerable guidewire available with a nominal diameter of 0.014 in (0.37 mm) and in nominal lengths of 182 cm. PT Graphix Magnet is available with a shapeable Straight Tip or a preformed "J"¿ Tip. The PT Graphix Magnet Guidewire features a 10 or 38 cm polymer sleeve, coated with ICE Hydrophilic Coating, jacketing the distal tapered core wire. The proximal section of the PT Graphix Magnet Guidewire is coated with a fluorinated polymer.

Publicado el 20 de abril de 2016 · Inicio del retiro: 28 de marzo de 2016

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Boston Scientific is voluntarily implementing a Medical Device Recall of one batch of PT Graphix Magnet J-TIP Guidewires. Boston Scientific has found that, while the batch is labeled as a J-tip model, the guidewires are straight and do not have the pre-formed J-tip. The most severe injury expected to occur is a limited procedure delay caused by the need to replace the guidewire with one that has

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Boston Scientific Corporation
Distribuidor
US nationwide in the states of ME, LA, and TN
Lote
Lot: 17971870
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States