Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1477-2024

stryker Infinity Resection Adjustment Block, REF 33600030, ankle arthroplasty

Publicado el 10 de abril de 2024 · Inicio del retiro: 6 de marzo de 2024

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Stryker has identified an issue that impacts specific lots of Infinity" Resection Guide Adjustment Blocks. The parts within these three lots were found to have been missing an internal screw in the finished instrument assembly. The missing screw is used to lockout the Medial/Lateral adjustment on the INFINITY Resection Adjustment Guide.

Qué hacer

Voluntary: Firm initiated · E-Mail

Detalles del retiro

Fabricante
Wright Medical Technology, Inc.
Distribuidor
US: TX, NE, and AUSTRALIA, BELGIUM, BRAZIL, CANADA, CHINA, COLOMBIA, FRANCE, GERMANY, ITALY, SPAIN, SWEDEN, SWITZERLAND, UK
Lote
UDI/DI 00889797004008, Lot Numbers: 2656950, 2762126, 2796094
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta