ECHELON 60mm Endoscopic Linear Cutter Reloads Black in Flexible Blister pack with Tyvek lid. The ECHELON and ECHELON FLEX families of Endoscopic Linear Cutters (articulating and straight) are sterile, single patient use instruments that simultaneously cut and staple tissue. There are six staggered rows of staples, three on either side of the cut line. The ECHELON 60 instruments have a staple line that is approximately 60 mm long and a cut line that is approximately 57 mm long. The shaft can rotate freely in both directions and an articulation mechanism on articulating instruments enables bending the distal portion of the shaft to facilitate lateral access of the operative site. The instruments are shipped without a reload and must be loaded prior to use. A staple retaining cap on the reload protects the staple leg points during shipping and transportation. The instruments'¿" lock-out feature is designed to prevent a used reload from being refired.
Publicado el 19 de junio de 2013 · Inicio del retiro: 13 de mayo de 2013
Resumen
FDA registra el retiro Z-1530-2013 de dispositivos médicos de Clase II, publicado el 19 de junio de 2013 e iniciado el 13 de mayo de 2013. Empresa responsable: Ethicon Endo-Surgery Inc; Estado: Finalizado; Cantidad reportada: 57,540 Units; Lote: The following lot numbers and associated expiration dates are affected by this recall: J4AY8Z 5/15/2013, J4C00J 6/15/2013, J4C87R 7/15/2013, J4CC1F 8/15/2013, J4CD2X 8/15/2013, J4CF3D 9/15/2013, J4CJ0E 9/15/2013, J4CJ78 9/15/2013, J4CK2P 9/15/2013, J4CM9X 10/15/2013, J4CR2H 10/15/2013, J4CT5H 11/15/2013, J4CU86 11/15/2013, J4CV1W 11/15/2013, J4CV5H 11/15/2013, K4C149 12/15/2013, K4C17W 12/15/2013, K4C33U 12/15/2013, K4C41C 12/15/2013, K4C600 1/15/2014, K4C64A 1/15/2014, 4C728 1/15/2014, K4C82K 1/15/2014, K4C88F 1/15/2014, K4C94W 1/15/2014, K4CC0L 2/15/2014, CC8L 2/15/2014, K4CD1E 2/15/2014, K4CF1P 2/15/2014, K4CF5D 2/15/2014, K4CG18 3/15/2014, K4CG4J 3/15/2014, K4CG8P 3/15/2014, K4CH44 3/15/2014, K4CK0W 3/15/2014,. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Ethicon Endo-Surgery is initiating a voluntary worldwide recall for ECHELON 60mm Black Reload (ECR60T) due to the potential for incomplete staple line formation from reload damage during the firing sequence.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Ethicon Endo-Surgery Inc
- Distribuidor
- Worldwide Distribution - US Nationwide all fifty states and including the countries of Canada, Belgium, India, Malaysia, Saudi Arabia, Brazil, Israel, Mexico, Singapore, Jordan, New Zealand, Taiwan, Chile, Korea, Puerto Rico, Turkey, Columbia, Kuwait, Qatar, United Arab Emirates, Egypt, Lebanon and Russia.
- Lote
- The following lot numbers and associated expiration dates are affected by this recall: J4AY8Z 5/15/2013, J4C00J 6/15/2013, J4C87R 7/15/2013, J4CC1F 8/15/2013, J4CD2X 8/15/2013, J4CF3D 9/15/2013, J4CJ0E 9/15/2013, J4CJ78 9/15/2013, J4CK2P 9/15/2013, J4CM9X 10/15/2013, J4CR2H 10/15/2013, J4CT5H 11/15/2013, J4CU86 11/15/2013, J4CV1W 11/15/2013, J4CV5H 11/15/2013, K4C149 12/15/2013, K4C17W 12/15/2013, K4C33U 12/15/2013, K4C41C 12/15/2013, K4C600 1/15/2014, K4C64A 1/15/2014, 4C728 1/15/2014, K4C82K 1/15/2014, K4C88F 1/15/2014, K4C94W 1/15/2014, K4CC0L 2/15/2014, CC8L 2/15/2014, K4CD1E 2/15/2014, K4CF1P 2/15/2014, K4CF5D 2/15/2014, K4CG18 3/15/2014, K4CG4J 3/15/2014, K4CG8P 3/15/2014, K4CH44 3/15/2014, K4CK0W 3/15/2014,
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAECHELON LINEAR Cutters Reload 80 mm Blue, intended for transection, resection, and/or creation of anastomoses,22 de octubre de 2025
- FDAENDOPATH ECHELON 45mm RELOADS WITH GRIPPING SURFACE TECHNOLOGY (GST), ECHELON Endopath Stapler, Endoscopic7 de febrero de 2024
- FDALigamax 5 mm Endoscopic Multiple Clip Applier18 de diciembre de 2019
- FDAECHELON FLEX Powered Plus ENDOPATH 60mm Stapler - 280mm shaft, Product Code PCEE60A6 de noviembre de 2019
- FDAUSHER® Support Catheter- Angled Tip WL130cm, RI 7F; REF USH07AT; PK1022-01A; Rx only, Non-Pyrogenic, Sterile;19 de junio de 2013
- FDAVIA Medical Collection Bag; The collection bag is an accessory waste bag used in conjunction with the sterile19 de junio de 2013
Preguntas sobre este retiro
¿ECHELON 60mm Endoscopic Linear Cutter Reloads Black in Flexible Blister pack with Tyvek lid. The ECHELON and ECHELON FLEX families of Endoscopic Linear Cutters (articulating and straight) are sterile, single patient use instruments that simultaneously cut and staple tissue. There are six staggered rows of staples, three on either side of the cut line. The ECHELON 60 instruments have a staple line that is approximately 60 mm long and a cut line that is approximately 57 mm long. The shaft can rotate freely in both directions and an articulation mechanism on articulating instruments enables bending the distal portion of the shaft to facilitate lateral access of the operative site. The instruments are shipped without a reload and must be loaded prior to use. A staple retaining cap on the reload protects the staple leg points during shipping and transportation. The instruments'¿" lock-out feature is designed to prevent a used reload from being refired. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1530-2013 el 19 de junio de 2013. Peligro reportado: Ethicon Endo-Surgery is initiating a voluntary worldwide recall for ECHELON 60mm Black Reload (ECR60T) due to the potential for incomplete staple line formation from reload damage during the firing sequence.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter