AW-M0001 SONARMED MONITOR. Used in conjunction with SonarMed Acoustic Sensors for airway monitoring.
Publicado el 24 de abril de 2024 · Inicio del retiro: 25 de marzo de 2024
Resumen
FDA registra el retiro Z-1535-2024 de dispositivos médicos de Clase I, publicado el 24 de abril de 2024 e iniciado el 25 de marzo de 2024. Empresa responsable: SonarMed Inc; Estado: En curso; Cantidad reportada: 145 units; Lote: Model No. AW-M0001; GTIN: 00851334007001; Serial No. AW0572, AW0573, AW0574, AW0577, AW0578, AW0579, AW0580, AW0581, AW0584, AW0585, AW0586, AW0588, AW0591, AW0593, AW0594, AW0600, AW0606, AW0607, AW0615, AW0616, AW0624, AW0626, AW0627, AW0648, AW0650, AW0651, AW0658, AW0664, AW0672, AW0674, AW0675, AW0678, AW0680, AW0681, AW0684, AW0685, AW0701, AW0702, AW0703, AW0704, AW0705, AW0706, AW0707, AW0708, AW0709, AW0710, AW0711, AW0712, AW0718, AW0719, AW0720, AW0721, AW0722, AW0723, AW0724, AW0725, AW0726, AW0727, AW0728, AW0729, AW0730, AW0731, AW0732, AW0733, AW0734, AW0735, AW0736, AW0737, AW0738, AW0739, AW0740, AW0741, AW0742, AW0743, AW0744, AW0745, AW0746, AW0747, AW0748, AW0749, AW0763, AW0764, AW0765, AW0766, AW0767, AW0768, AW0769, AW0770, AW0771, AW0772, AW0773, AW0774, AW0775, AW0776, AW0777, AW0778, AW0779, AW0780, AW0781, AW0782, AW0783, AW0784, AW0785, AW0786, AW0787, AW0788, AW0789, AW0790, AW0791, AW0792, AW0793, AW0794, AW0795, AW0796, AW0797, AW0798, AW0799, AW0800, AW0804, AW0805, AW0814, AW0817, AW0818, AW0819, AW0820, AW0821, AW0825, AW0826, AW0827, AW0828, AW0830, AW0831, AW0832, AW0833, AW0834, AW0835, AW0836, AW0837, AW0838, AW0839, AW0840, AW0841, AW0842, AW0854, AW0855. La Clase I indica una probabilidad razonable de consecuencias graves para la salud o muerte.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Software anomaly resulted in failure to detect a partial obstruction in 2.5 mm sensors and up to 3mm distal to the sensor.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- SonarMed Inc
- Distribuidor
- _AK, AZ, CA, FL, IL, KY, MD, MO, NC, NJ, NY, OH, OR, PA, SC, TN, TX, VA, & WA_
- Lote
- Model No. AW-M0001; GTIN: 00851334007001; Serial No. AW0572, AW0573, AW0574, AW0577, AW0578, AW0579, AW0580, AW0581, AW0584, AW0585, AW0586, AW0588, AW0591, AW0593, AW0594, AW0600, AW0606, AW0607, AW0615, AW0616, AW0624, AW0626, AW0627, AW0648, AW0650, AW0651, AW0658, AW0664, AW0672, AW0674, AW0675, AW0678, AW0680, AW0681, AW0684, AW0685, AW0701, AW0702, AW0703, AW0704, AW0705, AW0706, AW0707, AW0708, AW0709, AW0710, AW0711, AW0712, AW0718, AW0719, AW0720, AW0721, AW0722, AW0723, AW0724, AW0725, AW0726, AW0727, AW0728, AW0729, AW0730, AW0731, AW0732, AW0733, AW0734, AW0735, AW0736, AW0737, AW0738, AW0739, AW0740, AW0741, AW0742, AW0743, AW0744, AW0745, AW0746, AW0747, AW0748, AW0749, AW0763, AW0764, AW0765, AW0766, AW0767, AW0768, AW0769, AW0770, AW0771, AW0772, AW0773, AW0774, AW0775, AW0776, AW0777, AW0778, AW0779, AW0780, AW0781, AW0782, AW0783, AW0784, AW0785, AW0786, AW0787, AW0788, AW0789, AW0790, AW0791, AW0792, AW0793, AW0794, AW0795, AW0796, AW0797, AW0798, AW0799, AW0800, AW0804, AW0805, AW0814, AW0817, AW0818, AW0819, AW0820, AW0821, AW0825, AW0826, AW0827, AW0828, AW0830, AW0831, AW0832, AW0833, AW0834, AW0835, AW0836, AW0837, AW0838, AW0839, AW0840, AW0841, AW0842, AW0854, AW0855.
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAAW-S035 NEO SONARMED SENSOR 3.5MM. Used in conjunction with SonarMed Monitor for airway monitoring.24 de abril de 2024
- FDAAW-S025 NEO SONARMED SENSOR 2.5MM. Used in conjunction with SonarMed Monitor for airway monitoring.24 de abril de 2024
- FDAAW-S030 NEO SONARMED SENSOR 3.0MM. Used in conjunction with SonarMed Monitor for airway monitoring.24 de abril de 2024
- FDAPOLYSIN (Polyglycolic Acid) suture, HSM9 5-0 BR PGA Part Number G503N24 de abril de 2024
- FDAMasimo Rad-G, Pulse Oximeter with temperature (W/Sensor), REF:9210, Rx Only,24 de abril de 2024
- FDASTRATAFIX Spiral MONOCRYL Plus Bidirectional Knotless Tissue Control Device: STRATAFIX SPIRAL MONOCRYL PLUS BI24 de abril de 2024
Preguntas sobre este retiro
¿AW-M0001 SONARMED MONITOR. Used in conjunction with SonarMed Acoustic Sensors for airway monitoring. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1535-2024 el 24 de abril de 2024. Peligro reportado: Software anomaly resulted in failure to detect a partial obstruction in 2.5 mm sensors and up to 3mm distal to the sensor.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter