Mariner RDX Polyaxial Head, REF: MX1-000020; Mariner RDX Extended Polyaxial Head, REF: MX1-000030; Mariner RDX Deformity Head, REF: MX1-000040; Mariner RDX Extended Deformity Head, REF: MX1-000050; Mariner RDX Trauma Head, REF: MX1-000060; Mariner RDX Extended Trauma Head. REF: MX1-000070, part of Mariner RDX Alpha Instruments and Implants, REF: AMRDX and Mariner RDX TD Alpha Implants, REF: AMRDXTD, which are pedicle screw systems.
Publicado el 1 de mayo de 2024 · Inicio del retiro: 15 de febrero de 2024
Resumen
FDA registra el retiro Z-1605-2024 de dispositivos médicos de Clase II, publicado el 1 de mayo de 2024 e iniciado el 15 de febrero de 2024. Empresa responsable: SEASPINE ORTHOPEDICS CORPORATION; Estado: En curso; Cantidad reportada: 1579 Screw Heads; Lote: REF/UDI-DI/Lots: MX1-000020/10889981289751/AW161073C, AW164798D, AW164799D, AW164800E, AW164801E; MX1-000030/10889981289768/AW164802F, AW164803G, AW157588F, AW161074F, AW163331F; MX1-000040/10889981289775/AW157589D, AW164806E; MX1-000050/10889981289782/AW164807G, AW164808G, AW164809H, AW157590G, AW172902H; MX1-000060/10889981289799/AW157591D, AW161076E, AW164811E; MX1-000070/10889981289805/AW163350G, AW164812G, AW157592G, AW161077G; AMRDX/10889981291266/AMRDX-001, AMRDX-002, AMRDX-003, AMRDX-004, AMRDX-005, AMRDX-008, AMRDX-009; AMRDXTD/10889981295332/AMRDXTD-003, AMRDXTD-004, AMRDXTD-005. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Screw heads, of pedicle screw systems, have thin lead-in thread that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- SEASPINE ORTHOPEDICS CORPORATION
- Distribuidor
- US Nationwide distribution in the states of MI, GA, VA, IN, FL.
- Lote
- REF/UDI-DI/Lots: MX1-000020/10889981289751/AW161073C, AW164798D, AW164799D, AW164800E, AW164801E; MX1-000030/10889981289768/AW164802F, AW164803G, AW157588F, AW161074F, AW163331F; MX1-000040/10889981289775/AW157589D, AW164806E; MX1-000050/10889981289782/AW164807G, AW164808G, AW164809H, AW157590G, AW172902H; MX1-000060/10889981289799/AW157591D, AW161076E, AW164811E; MX1-000070/10889981289805/AW163350G, AW164812G, AW157592G, AW161077G; AMRDX/10889981291266/AMRDX-001, AMRDX-002, AMRDX-003, AMRDX-004, AMRDX-005, AMRDX-008, AMRDX-009; AMRDXTD/10889981295332/AMRDXTD-003, AMRDXTD-004, AMRDXTD-005
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDASeaSpine Mariner Deformity Product Name: Extended Polyaxial Head, Attached Gimbal, 12mm REF: MD1-300010-1230 de septiembre de 2026
- FDASeaSpine Polyaxial Head REF Number: 41-301029 de julio de 2026
- FDABrand Name: Daytona Small Stature Product Name: CoCr LockingCap 4.5mm, Small Stature, T25, BT Model/Catalog1 de enero de 2025
- FDABrand Name: Meridian Angled Awl Fixed Product Name: Angled Awl Fixed Model/Catalog Number: RA2-003015 Product24 de julio de 2024
- FDARedux Electrolyte Creme, Product Ref. 66-04. Highly conductive electrolyte creme.1 de mayo de 2024
- FDAVarious Alcon Custom Paks Surgical Procedure Pack containing Oasis 27Ga Chang Hydrodissection Cannulas. These1 de mayo de 2024
Preguntas sobre este retiro
¿Mariner RDX Polyaxial Head, REF: MX1-000020; Mariner RDX Extended Polyaxial Head, REF: MX1-000030; Mariner RDX Deformity Head, REF: MX1-000040; Mariner RDX Extended Deformity Head, REF: MX1-000050; Mariner RDX Trauma Head, REF: MX1-000060; Mariner RDX Extended Trauma Head. REF: MX1-000070, part of Mariner RDX Alpha Instruments and Implants, REF: AMRDX and Mariner RDX TD Alpha Implants, REF: AMRDXTD, which are pedicle screw systems. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1605-2024 el 1 de mayo de 2024. Peligro reportado: Screw heads, of pedicle screw systems, have thin lead-in thread that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter