Dimension Vista¿ CSA and CSAE The Cyclosporine method is an in vitro diagnostic test for the quantitative measurement of cyclosporine A (CSA) in human whole blood on the Dimension Vista¿ System. Measurements of CSA are used as an aid in the management of heart, liver and kidney transplant patients.
Publicado el 28 de mayo de 2014 · Inicio del retiro: 19 de marzo de 2014
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
A complaint was received of Dimension Vista¿ CSA and CSAE under-recovery of cyclosporine when patient samples are run from the Small Sample Container. The under-recovery that has been observed is 18 to 44%. In March 2014, Siemens recalled due to the complaint received of Dimension Vista¿ CSA and CSAE reporting under-recovery of cyclosporine when patient samples are run from the Small Sample C
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Siemens Healthcare Diagnostics, Inc.
- Distribuidor
- Worldwide Distribution - USA (nationwide) and Internationally to Canada, Belgium, Switzerland, Germany, Spain, France, Great Britain, Italy, Saudi Arabia, Australia, and Japan.
- Lote
- CSA 13114BA, exp. 2014-04-24; CSA 13157BA, exp. 2014-06-06; CSA 13246BB, exp. 2014-09-03; CSA 13317BA, exp. 2014-11-13; CSA 14008BB, exp. 2015-01-08; CSA 14050BB, exp. 2015-02-19; CSAE 13157BC, exp. 2014-06-06; CSAE 13249BD, exp. 2014-09-06.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States