FDAFDA · Alimentos y medicamentos·N.º oficial Z-1709-2016
3.2 mm Three-fluted drill bit, sterile, Catalog ID # 03.010.060S Product Usage: Usage: The 3.2 mm Three-fluted Drill Bits are instruments that can be used with the following systems: Expert Lateral Femoral Nail (part number 03.010.060S only), Expert Tibial Nail, Expert Humeral Nailing, Multiloc Humeral Nailing System, Suprapatellar Instrumentation for Expert Tibial Nail, and Adolescent Lateral Femoral Nails (ALFN).
Publicado el 1 de junio de 2016 · Inicio del retiro: 18 de abril de 2016
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The sterile packaging may be compromised. A protective cap is to be secured on the tip of the drill bit. If the protective cap is not secured, the drill bit may puncture the sterile barrier.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Synthes (USA) Products LLC
- Distribuidor
- US Nationwide distribution in the states of: AE, AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX,UT, VA, VT, WA, WI
- Lote
- Catalog ID 03.010.060S Lot # 5558809, 6204324, 6813586, 7648837, 5559839, 6214366, 6818438, 7648838, 5575915, 6235676, 6827807, 7657752, 5579717, 6240287, 6830470, 7657753, 5582298, 6247514, 6835634, 7660380, 5585788, 6249834, 6841962, 7661528 5594237, 6269336, 6851101, 7661529, 5594238, 6276953, 6857546, 7674414, 5616439, 6279756, 6860704, 7687319, 5633067, 6279757, 6870563, 7693164, 5656914, 6302746, 6873394, 7716604, 5672506, 6313568, 6873396, 7731690 5682043, 6313569, 6898040, 7739027, 5692087, 6325420, 6904377, 7753584, 5702206, 6331923, 6908328, 7766654, 5711966, 6334954, 6919278, 7774127, 5724637, 6339936, 6925232, 7774128, 5727691, 6348925, 6931055, 7794479 5732158, 6355191, 6939080, 7817283, 5741332, 6368986, 6949503, 7830606, 5748361, 6375334, 6961056, 7835732, 5760640, 6382329, 6967764, 7848333, 5772367, 6398746, 6975982, 7869029, 5780654, 6413506, 6984973, 7870629 5782800, 6420887, 6996252, 7894923, 5782801, 6433683, 7004816, 7899079, 5782802, 6433684, 7013143, 7903010, 5794181, 6438144, 7019693, 7908522, 5816780, 6449832, 7029024, 7918012, 5818769, 6458079, 7036198, 7919639 5818770, 6464443, 7052563, 7924989, 5829138, 6475737, 7064072, 7941743, 5847359, 6480417, 7070681, 7947278, 5860702, 6486376, 7084937, 7963626, 5861834, 6488940, 7275546, 7981633, 5861836, 6505811, 7304754, 7993351 5870708, 6509276, 7349355, 9808655, 6010554, 6528479, 7359254, 9820617, 6031951, 6540263, 7371663, 9825386, 6034093, 6549146, 7382623, 9826363, 6034128, 6557658, 7400014, 9831036, 6051790, 6563112, 7400015, 9841499 6052762, 6572199, 7400016, 9855040, 6055036, 6577880, 7416878, 9859105, 6066429, 6586607, 7422660, 9864677, 6069150, 6593553, 7441691, 9869742, 6069151, 6598206, 7444103, 9878216, 6074684, 6603184, 7460590, 9886333 6080184, 6610704, 7465210, 9904888, 6081960, 6621162, 7476937, 9911126, 6091446, 6636693, 7485616, 9915461, 6092777, 6655352, 7497778, 9924572, 6109266, 6674053, 7499940, 9932287, 6121870, 6686555, 7512119, 9939768 6124816, 6707678, 7520264, 9959608, 6130838, 6721722, 7533286, 9961574, 6133623, 6733633, 7544212, 9970453, 6136836, 6752258, 7554690, 9985425, 6141455, 6761019, 7563621, H023093, 6153365, 6766654, 7636777, H024812 6163583, 6777654, 7636778, H034525, 6186051, 6786129, 7636779, H039109, 6186052, 6796018, 7648835, 6187891, 6796019, 7648836
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States