illumipro-10 Incubator/Reader, Meridian Bioscience, Inc. The illumipro-10 incubator/reader is designed to be used with illumigene Molecular Assays. The device incubates illumigene Test Devices and reads endpoint reactions.
Publicado el 9 de julio de 2014 · Inicio del retiro: 22 de abril de 2014
Resumen
FDA registra el retiro Z-1941-2014 de dispositivos médicos de Clase II, publicado el 9 de julio de 2014 e iniciado el 22 de abril de 2014. Empresa responsable: Meridian Bioscience Inc; Estado: Finalizado; Cantidad reportada: 35 instruments; Lote: Serial Numbers: IP2669, IP2670, IP2671, IP2672, IP2673, IP2674, IP2675, IP2676, IP2677, IP2678, IP2680, IP2682, IP2683, IP2684, IP2685, IP2686, IP2687, IP2688, IP2689, IP2690, IP2691, IP2692, IP2693, IP2694, IP2695, IP2696, IP2697, IP2698, IP2699, IP2700, IP2701, IP2702, IP2703, IP2704, IP2705, IP2706. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Illumipro-10 instruments may have Block B chambers operating at an incorrect temperature as a result of incorrect temperature calibration.
Qué hacer
Voluntary: Firm initiated · Telephone
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Meridian Bioscience Inc
- Distribuidor
- Worldwide Distribution - USA including AL, AZ, CA, CO, FL, IA, IN, LA, MA, ME, MI, MN, NJ, NY, OH, PA, and WV. Internationally to Australia.
- Lote
- Serial Numbers: IP2669, IP2670, IP2671, IP2672, IP2673, IP2674, IP2675, IP2676, IP2677, IP2678, IP2680, IP2682, IP2683, IP2684, IP2685, IP2686, IP2687, IP2688, IP2689, IP2690, IP2691, IP2692, IP2693, IP2694, IP2695, IP2696, IP2697, IP2698, IP2699, IP2700, IP2701, IP2702, IP2703, IP2704, IP2705, IP2706.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDARevogene SARS-CoV-2_IVD real-time RT-PCR test intended for the qualitative detection of RNA from SARSCoV-21 de febrero de 2023
- FDARevogene, For In Vitro Diagnostic use in performing nucleic acid testing in clinical laboratories. Meridian6 de abril de 2022
- FDAPREMIER Cryptococcal Antigen, Catalog Number 602096 Product Usage: The PREMIER Cryptococcal Antigen enzyme23 de enero de 2019
- FDAStirling Ultracold Portable Shuttle, ULT-25. Storage for cold-chain management, short-term or long-term9 de julio de 2014
- FDAStride Femoral Component Size 2, PFSI-00053, Right Lateral /Left Medial. Unicondylar Knee prosthesis.9 de julio de 2014
- FDANexGen Complete Knee Solution, MIS Total Knee Procedure, Stemmed Tibial Component, Precoat, Fixed Bearing,9 de julio de 2014
Preguntas sobre este retiro
¿illumipro-10 Incubator/Reader, Meridian Bioscience, Inc. The illumipro-10 incubator/reader is designed to be used with illumigene Molecular Assays. The device incubates illumigene Test Devices and reads endpoint reactions. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1941-2014 el 9 de julio de 2014. Peligro reportado: Illumipro-10 instruments may have Block B chambers operating at an incorrect temperature as a result of incorrect temperature calibration.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Telephone