FDAFDA · Alimentos y medicamentos·N.º oficial Z-1952-2015
Milestone Wand STA Handpiece for the delivery of local anesthesia to dental tissue. Model Number STA-5050-305. The Wand STA handpiece is a single use syringe used in conjunction with the CompuDent STA computer controlled dental injection system.
Publicado el 8 de julio de 2015 · Inicio del retiro: 11 de febrero de 2015
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Wand Handpiece was mislabeled as STA handpiece. The Wand handpiece cannot be installed in the STA instrument.
Qué hacer
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Detalles del retiro
- Fabricante
- Milestone Scientific, Inc.
- Distribuidor
- Worldwide distribution. US in FL, IL, NY WA and Puerto Rico, Israel, Croatia, Italy, Australia, Germany, Colombia, UK, Sweden, and UAE.
- Lote
- Lot 140103
- Clasificación de riesgo
- Class III
- Área (FDA)
- device
- País
- United States