Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2456-2026

TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816

Publicado el 24 de junio de 2026 · Inicio del retiro: 16 de marzo de 2026

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Humeral nail due to a manufacturing error is missing distal holes and if implanted may lead to revision surgery but, if the missing holes are identified before implantation, an equivalent backup can be used, or a bent nail of the same size, which would require additional surgical steps to address soft tissue/bone preparation, which could lead to a surgical delay of greater than 30 minutes.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Smith & Nephew, Inc.
Distribuidor
International distribution to the countries of BE, DE, IT.
Lote
UDI-DI: 03596010514394. Lot/Batch: 25BM03616
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta