ADVIA Chemistry XPT, SMN 10723034, IVD. --- This issue affects only the ADVIA Chemistry Hemoglobin A1c_3 Automated Pretreatment (A1c_3) results when HbA1c values are reported in International Federation of Clinical Chemistry (IFCC) equivalent units (HbA1cR). The ADVIA Chemistry XPT System is an automated, clinical chemistry analyzer that runs tests on serum, plasma, urine, or cerebral spinal fluid in random access and batch modes at a throughput rate of both 1800 photometric tests per hour and 600 electrolyte (ISE) tests per hour
Publicado el 7 de septiembre de 2016 · Inicio del retiro: 20 de julio de 2016
Resumen
FDA registra el retiro Z-2704-2016 de dispositivos médicos de Clase II, publicado el 7 de septiembre de 2016 e iniciado el 20 de julio de 2016. Empresa responsable: Siemens Healthcare Diagnostics, Inc.; Estado: Finalizado; Cantidad reportada: Domestic: 32 units; Foreign: 197 units; Lote: CA1275002940294 and lower. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Siemens identified an issue with the ratio equation provided on the ADVIA Chemistry XPT System Software Test Definition (TDef) Version 1.0 disks (SMN: 11127343 and 11222123). This issue affects only the ADVIA Chemistry Hemoglobin A1c_3 Automated Pretreatment (A1c_3) results when HbA1c values are reported in International Federation of Clinical Chemistry (IFCC) equivalent units (HbA1cR). The ratio equation used to calculate ADVIA Chemistry Hemoglobin A1c_3 Automated Pretreatment (A1c_3) results in IFCC units contains an error. The error results in falsely depressed HbA1cR results.
Qué hacer
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Siemens Healthcare Diagnostics, Inc.
- Distribuidor
- Worldwide Distribution - US to MA, NY, MI, IL, TX, CO, AZ, CA, WA and Foreign distribution to the following countries: Australia, Austria, Bahrain, Brazil, China, Denmark, Egypt, Finland, France, Germany, Greece, Hungary, Italy, Malaysia, Netherlands, New Zealand, Norway, Portugal, Portugal, Republic Korea, Singapore, Slovakia, Spain, Sweden, Turkey, United Kingdom, and Vietnam.
- Lote
- CA1275002940294 and lower
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAAtellica CH Revised C-Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223;7 de octubre de 2026
- FDABeckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mL26 de agosto de 2026
- FDASyva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL, 1x14mL26 de agosto de 2026
- FDASyva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219UL12 de agosto de 2026
- FDAUniCel DxH Slidemaker Stainer Coulter Cellular Analysis System, Catalog No. 775222 UniCel DxH Slidemaker7 de septiembre de 2016
- FDADouble Pump RF Patient Cassette; Product Number: 72204055. Intended to provide fluid distension and irrigation7 de septiembre de 2016
Preguntas sobre este retiro
¿ADVIA Chemistry XPT, SMN 10723034, IVD. --- This issue affects only the ADVIA Chemistry Hemoglobin A1c_3 Automated Pretreatment (A1c_3) results when HbA1c values are reported in International Federation of Clinical Chemistry (IFCC) equivalent units (HbA1cR). The ADVIA Chemistry XPT System is an automated, clinical chemistry analyzer that runs tests on serum, plasma, urine, or cerebral spinal fluid in random access and batch modes at a throughput rate of both 1800 photometric tests per hour and 600 electrolyte (ISE) tests per hour fue retirado del mercado?
Sí. FDA publicó el retiro Z-2704-2016 el 7 de septiembre de 2016. Peligro reportado: Siemens identified an issue with the ratio equation provided on the ADVIA Chemistry XPT System Software Test Definition (TDef) Version 1.0 disks (SMN: 11127343 and 11222123). This issue affects only the ADVIA Chemistry Hemoglobin A1c_3 Automated Pretreatment (A1c_3) results when HbA1c values are reported in International Federation of Clinical Chemistry (IFCC) equivalent units (HbA1cR). The ratio equation used to calculate ADVIA Chemistry Hemoglobin A1c_3 Automated Pretreatment (A1c_3) results in IFCC units contains an error. The error results in falsely depressed HbA1cR results.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit