Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2775-2026

SnapshotGLO Wound Imaging Device, Model KB100, KB100AA.

Publicado el 29 de julio de 2026 · Inicio del retiro: 30 de junio de 2026

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The software controlling the wound imaging device's battery charging circuit was not designed to recognize and recover a battery that has fully discharged into a low-power "shipping mode." When this occurs, an affected device will not resume charging when reconnected to power and becomes temporarily unusable/ unavailable for wound imaging until corrected.

Qué hacer

Voluntary: Firm initiated

Detalles del retiro

Fabricante
KENT IMAGING, INC.
Distribuidor
US Nationwide distribution in the states of PA, IA, FL, TX, GA, NV, CA, UT, IN, DC, KS, IL, TN, AR, AZ.
Lote
UDI-DI: B376KB1000, Software Revisions v4.2.0 and v4.1.3
Clasificación de riesgo
Class II
Área (FDA)
device
País
Canada