Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0160-2015

Terumo Custom Cardioplegia Delivery Set- Catalog Number:165720 and Catalog Number: 6375 Product Usage: Cardioplegia Delivery System that is designed to delivery blood/solution to the patient undergoing cardiopulmonary bypass surgery. The set is designed to deliver blood/solutions at a 4: I ratio. Includes a Conducer Heat Exchanger and a MP4 Temperature & Pressure Module. Additionally, the set is constructed with a female luer connector that is the subject of the correction/removal activity.

Publicado el 12 de noviembre de 2014 · Inicio del retiro: 1 de marzo de 2011

Resumen

FDA registra el retiro Z-0160-2015 de dispositivos médicos de Clase II, publicado el 12 de noviembre de 2014 e iniciado el 1 de marzo de 2011. Empresa responsable: Terumo Cardiovascular Systems Corp; Estado: Finalizado; Cantidad reportada: Catalog: 16579 ; Catalog 6375: 3256 units; Lote: Catalog Number: 165720 Lot Numbers: MDI5, MD15B, MD22, ME19, MP13. Manufacturing dates March 15, 20 I 0 through December 13, 20 I 0. Catalog Number: 6375 Lot Numbers: MD15, ME19, MPI3 Manufacturing dates March 15, 2010 through December 13, 20 I 0. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Defect in the luer connector can compromise the seal on the female side of the connector. and the compromised seal can introduce air in the table line

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Terumo Cardiovascular Systems Corp
Distribuidor
Worldwide Distribution: US (nationwide) and the countries of Canada, Brazil, Hong Kong, Korea, and Thailand.
Lote
Catalog Number: 165720 Lot Numbers: MDI5, MD15B, MD22, ME19, MP13. Manufacturing dates March 15, 20 I 0 through December 13, 20 I 0. Catalog Number: 6375 Lot Numbers: MD15, ME19, MPI3 Manufacturing dates March 15, 2010 through December 13, 20 I 0
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Terumo Custom Cardioplegia Delivery Set- Catalog Number:165720 and Catalog Number: 6375 Product Usage: Cardioplegia Delivery System that is designed to delivery blood/solution to the patient undergoing cardiopulmonary bypass surgery. The set is designed to deliver blood/solutions at a 4: I ratio. Includes a Conducer Heat Exchanger and a MP4 Temperature & Pressure Module. Additionally, the set is constructed with a female luer connector that is the subject of the correction/removal activity. fue retirado del mercado?

Sí. FDA publicó el retiro Z-0160-2015 el 12 de noviembre de 2014. Peligro reportado: Defect in the luer connector can compromise the seal on the female side of the connector. and the compromised seal can introduce air in the table line

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter