4.5mm PLEO Osteotomy Guide. OrthoPediatrics Corp. markets a pediatric, extension osteotomy bone plate system. This system is marketed under the brand name OrthoPediatrics PediLock Extension Osteotomy Plate ( PLEO ). The product included in this recall is a Class I, 4.5mm PLEO Osteotomy Guide Instrument, common name: Guide, Surgical, Instrument. The PediLoc® Extension Osteotomy Plate is part of the OrthoPediatrics~PediLoc® system. The PediLoc® Extension Osteotomy Plate consists of 3.5mm and 4.5mm plates, left and right side specific. The PediLoc® Extension Osteotomy Plate is designed to fit the anatomy of the distal femur in children and adolescents, above the distal femoral physis. The distal section of the plate is designed so that three locking screws can be placed in the distal bone segment at an angle parallel to the distal femoral physis. The proximal portion of the plate is designed to accept locking or non-locking screws. The subject instrument of this recall is an adjunct instrument for the convenience of some surgeons who choose to use it. This instrument is not illustrated in the Surgical Technique, nor required to perform the procedure.

Publicado el 28 de noviembre de 2012 · Inicio del retiro: 18 de octubre de 2012

Resumen

FDA registra el retiro Z-0432-2013 de dispositivos médicos de Clase II, publicado el 28 de noviembre de 2012 e iniciado el 18 de octubre de 2012. Empresa responsable: OrthoPediatrics Corp; Estado: Finalizado; Cantidad reportada: 24; Lote: Part number: 01-1050-0066 and 01-1050-0067 Lot numbers: 7861001 and 7861101 only. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Complaint received from sales representative that the drill guide tubes would not screw into replenishment instrument. Upon investigation, it has been determined that the device was not manufactured to design specifications and will not function as intended.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
OrthoPediatrics Corp
Distribuidor
Nationwide Distribution-including the states of OH, NC, CO, FL, IN, MI, and TX.
Lote
Part number: 01-1050-0066 and 01-1050-0067 Lot numbers: 7861001 and 7861101 only
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿4.5mm PLEO Osteotomy Guide. OrthoPediatrics Corp. markets a pediatric, extension osteotomy bone plate system. This system is marketed under the brand name OrthoPediatrics PediLock Extension Osteotomy Plate ( PLEO ). The product included in this recall is a Class I, 4.5mm PLEO Osteotomy Guide Instrument, common name: Guide, Surgical, Instrument. The PediLoc® Extension Osteotomy Plate is part of the OrthoPediatrics~PediLoc® system. The PediLoc® Extension Osteotomy Plate consists of 3.5mm and 4.5mm plates, left and right side specific. The PediLoc® Extension Osteotomy Plate is designed to fit the anatomy of the distal femur in children and adolescents, above the distal femoral physis. The distal section of the plate is designed so that three locking screws can be placed in the distal bone segment at an angle parallel to the distal femoral physis. The proximal portion of the plate is designed to accept locking or non-locking screws. The subject instrument of this recall is an adjunct instrument for the convenience of some surgeons who choose to use it. This instrument is not illustrated in the Surgical Technique, nor required to perform the procedure. fue retirado del mercado?

Sí. FDA publicó el retiro Z-0432-2013 el 28 de noviembre de 2012. Peligro reportado: Complaint received from sales representative that the drill guide tubes would not screw into replenishment instrument. Upon investigation, it has been determined that the device was not manufactured to design specifications and will not function as intended.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter