ACCU-CHEK Tender II 13/110 10 + 10 pieces Product Usage: Intended for the infusion and/or injection of fluids into the body below the surface of the skin. The indwelling catheter can be inserted independently from the infusion catheter and can be accessed for the injections through the injection port. The indwelling catheter can also be securely attached to the infusion catheter by means of a proprietary click-lock connector for the infusion of drugs subcutaneously. The infusion set can be detached from the indwelling catheter, and the catheter capped to allow freedom from the infusion set and pump for showers, athletics or other activities.
Publicado el 7 de enero de 2015 · Inicio del retiro: 20 de noviembre de 2014
Resumen
FDA registra el retiro Z-0891-2015 de dispositivos médicos de Clase II, publicado el 7 de enero de 2015 e iniciado el 20 de noviembre de 2014. Empresa responsable: Roche Diagnostics Operations, Inc.; Estado: Finalizado; Cantidad reportada: Total of 13,041 boxes; Lote: Catalog: 04541456001 and lot: 5052067. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Roche Diabetes Care was informed by UnoMedical, the Manufacturer of ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets, about potential for the the tubing of the infusion set to become detached at the connect/disconnect location on the ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets. If tubing detachment occurs, insulin delivery is interrupted and the pump will not alarm to notify
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Roche Diagnostics Operations, Inc.
- Distribuidor
- US Nationwide Distribution
- Lote
- Catalog: 04541456001 and lot: 5052067
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAcobas HCYS, Homocysteine Enzymatic Assay, Material Number REF 06542921190, for use on cobas c 701/702, in27 de noviembre de 2024
- FDACreatine Kinase (CK) used on cobas c 311/501/502 and COBAS INTEGRA 400 plus-In vitro test for the quantitative2 de octubre de 2024
- FDAAlbumin Gen.2, Catalog 0516686119021 de octubre de 2020
- FDAElecsys Anti-HAV IgM assay-for the in vitro qualitative detection of IgM antibodies to hepatitis A virus18 de diciembre de 2019
- FDAACCU-CHEK Tender I 17/80 10 pieces Product Usage: Intended for the infusion and/or injection of fluids into7 de enero de 2015
- FDACryogenic probe for cardiac ablation surgery, packaged individually in Tyvek pouches Product Usage: Cryogenic7 de enero de 2015
Preguntas sobre este retiro
¿ACCU-CHEK Tender II 13/110 10 + 10 pieces Product Usage: Intended for the infusion and/or injection of fluids into the body below the surface of the skin. The indwelling catheter can be inserted independently from the infusion catheter and can be accessed for the injections through the injection port. The indwelling catheter can also be securely attached to the infusion catheter by means of a proprietary click-lock connector for the infusion of drugs subcutaneously. The infusion set can be detached from the indwelling catheter, and the catheter capped to allow freedom from the infusion set and pump for showers, athletics or other activities. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0891-2015 el 7 de enero de 2015. Peligro reportado: Roche Diabetes Care was informed by UnoMedical, the Manufacturer of ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets, about potential for the the tubing of the infusion set to become detached at the connect/disconnect location on the ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets. If tubing detachment occurs, insulin delivery is interrupted and the pump will not alarm to notify
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter