FDAFDA · Alimentos y medicamentos·N.º oficial Z-0931-2018

Proteus 235, Pencil Beam Scanning version PTS-6.4.11.X The Proton Therapy System - Proteus 235 (brand names: Proteus Plus and Proteus ONE) is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. The PTS may include a fixed small beam treatment room dedicated to the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation localized to the head and neck.

Publicado el 14 de marzo de 2018 · Inicio del retiro: 19 de octubre de 2017

Resumen

FDA registra el retiro Z-0931-2018 de dispositivos médicos de Clase II, publicado el 14 de marzo de 2018 e iniciado el 19 de octubre de 2017. Empresa responsable: Ion Beam Applications S.A.; Estado: Finalizado; Cantidad reportada: 3; Lote: PAT.003 (KR), PAT.006 (US). La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Range for the same energy could be different depending if the layer is the first layer of the irradiation plan or not.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Ion Beam Applications S.A.
Distribuidor
US Distribution to the state of :FL
Lote
PAT.003 (KR), PAT.006 (US)
Clasificación de riesgo
Class II
Área (FDA)
device
País
Belgium

Retiros relacionados

Preguntas sobre este retiro

¿Proteus 235, Pencil Beam Scanning version PTS-6.4.11.X The Proton Therapy System - Proteus 235 (brand names: Proteus Plus and Proteus ONE) is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. The PTS may include a fixed small beam treatment room dedicated to the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation localized to the head and neck. fue retirado del mercado?

Sí. FDA publicó el retiro Z-0931-2018 el 14 de marzo de 2018. Peligro reportado: Range for the same energy could be different depending if the layer is the first layer of the irradiation plan or not.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter