Drill Guide - Chameleon FacetFuse Fixation System, Model: 11-00028 Rev. A FacetFuse Drill Guides were designed to provide a concentric pathway for the drills and taps to access the bone while protecting the soft tissue from the cutting geometry of the drills/taps. In addition the drill guide is used as a reference to measure depth that the drills and taps are protruding into bone, which is measured by depth markings at the proximal end of the drills/taps relative to the proximal end of the drill guide. These devices were used during implantation of spinal intervertebral body fusion devices intended for intervertebral body fusion of the spine of skeletally mature patients.
Publicado el 14 de agosto de 2013 · Inicio del retiro: 9 de noviembre de 2012
Resumen
FDA registra el retiro Z-1871-2013 de dispositivos médicos de Clase II, publicado el 14 de agosto de 2013 e iniciado el 9 de noviembre de 2012. Empresa responsable: SpineFrontier, Inc.; Estado: Finalizado; Cantidad reportada: 13 drill guides; Lote: 2520301, 2520307. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Prior to use in surgery, the Drill Guides were presenting noticeable degrees of denting and/or bending. As a result, drills and taps could not be directed down guides during surgery due to mechanical interference.
Qué hacer
Voluntary: Firm initiated · E-Mail
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- SpineFrontier, Inc.
- Distribuidor
- US Distribution including Puerto Rico and the states of CO, KS, MA and TX.
- Lote
- 2520301, 2520307
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAA-CIFT SoloFuse Fixed Angle Driver, Model Number 13-32101-01 (Component Part Numbers 14-32107-01 and 14-32106)24 de julio de 2019
- FDAlnline Persuader- PedFuse Pedicle Screw System; Part Number: 11-8016421 de enero de 2015
- FDASpineFrontier S-Lift Lock Shaft Inserter, an instrument used with S-LIFT Intervertebral Body Fusion Device in14 de mayo de 2014
- FDAArena-L Trial Handle; Part: 11-81002 The trial heads are smooth plates that mimic the footprint of the22 de enero de 2014
- FDAList No. 14219-28; Y-TYPE BLOOD SET; 200 Micron Filter, Cylinder Pump, CLAVE Y-Site, Non-Vented, 80 Inch,14 de agosto de 2013
- FDANon-sterile Viscot Surgical Skin Markers Part Number 1437-100 Non-sterile Viscot Part Number 143-100 Surgical14 de agosto de 2013
Preguntas sobre este retiro
¿Drill Guide - Chameleon FacetFuse Fixation System, Model: 11-00028 Rev. A FacetFuse Drill Guides were designed to provide a concentric pathway for the drills and taps to access the bone while protecting the soft tissue from the cutting geometry of the drills/taps. In addition the drill guide is used as a reference to measure depth that the drills and taps are protruding into bone, which is measured by depth markings at the proximal end of the drills/taps relative to the proximal end of the drill guide. These devices were used during implantation of spinal intervertebral body fusion devices intended for intervertebral body fusion of the spine of skeletally mature patients. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1871-2013 el 14 de agosto de 2013. Peligro reportado: Prior to use in surgery, the Drill Guides were presenting noticeable degrees of denting and/or bending. As a result, drills and taps could not be directed down guides during surgery due to mechanical interference.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · E-Mail