Perouse Flamingo, Model 0218TA and 0218TB intended for use during cardiovascular procedure to create, maintain and monitor pressure in the balloon catheter.
Publicado el 30 de julio de 2014 · Inicio del retiro: 10 de mayo de 2012
Resumen
FDA registra el retiro Z-2089-2014 de dispositivos médicos de Clase II, publicado el 30 de julio de 2014 e iniciado el 10 de mayo de 2012. Empresa responsable: Perouse Medical; Estado: Finalizado; Cantidad reportada: 2,496 units; Lote: Model 0218TA and 0218TB. Manufacturing Lot Numbers: 12031275 and 12031242. EXP.: 02/2017. Batches: 12031526, 12031249, 12031527, 12031248, 12041183, 12041186, 12031244, 12031461, 12031460, 12021404, 12031462, 12031245, 12021405, 12031405, 12031246, 12021401, 12041344, 12031237, 12031241, 12021403, 12041345, 12041184, 12031497, 12031240, 12021194, 12031524, 12031480, 12041181, 12031275, 12031242, 12021450, 12021449, 12041185, 12031525, 12031247, 12021402, 12021448. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Perouse Medical in France initiated a voluntary recall of certain batches of Perouse FLAMINGO Inflations Devices due to a potential water-tightness default. The potential default could cause an air entry when the plunger is positioned around 20 mL.
Qué hacer
Voluntary: Firm initiated · E-Mail
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Perouse Medical
- Distribuidor
- US.
- Lote
- Model 0218TA and 0218TB. Manufacturing Lot Numbers: 12031275 and 12031242. EXP.: 02/2017. Batches: 12031526, 12031249, 12031527, 12031248, 12041183, 12041186, 12031244, 12031461, 12031460, 12021404, 12031462, 12031245, 12021405, 12031405, 12031246, 12021401, 12041344, 12031237, 12031241, 12021403, 12041345, 12041184, 12031497, 12031240, 12021194, 12031524, 12031480, 12041181, 12031275, 12031242, 12021450, 12021449, 12041185, 12031525, 12031247, 12021402, 12021448.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- France
Retiros relacionados
- FDAHemochron Jr. Whole Blood Microcoagulation System APTT Cuvette. For in vitro diagnostic coagulation test.30 de julio de 2014
- FDAZoll Circulation LifeBand Single Pack (accessory), and Nickel-metal hydride (NIMH) Battery. The LifeBand is a30 de julio de 2014
- FDASynthes Guide Blocks for the 2 Column Plate 6 Hole Head (Volar Distal Radius Plate System). Orthopedic30 de julio de 2014
- FDAMK5 NX Controller--Wheelchair controller30 de julio de 2014
- FDANOVATION Femoral Stem, Tapered Extended offset, plasma- coated, press fit. Catalog number 160-01-13 intended30 de julio de 2014
- FDASolanas 2.1mm FIXED DEPTH DRILL, 10mm, Part No. 63928-10. Component of the SOLANAS III Posterior Stabilization30 de julio de 2014
Preguntas sobre este retiro
¿Perouse Flamingo, Model 0218TA and 0218TB intended for use during cardiovascular procedure to create, maintain and monitor pressure in the balloon catheter. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2089-2014 el 30 de julio de 2014. Peligro reportado: Perouse Medical in France initiated a voluntary recall of certain batches of Perouse FLAMINGO Inflations Devices due to a potential water-tightness default. The potential default could cause an air entry when the plunger is positioned around 20 mL.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · E-Mail