Tibila Insert UC, Balanced Knee System, REF163-3YYY Product Usage: UHMWPE component used in TKA procedures to serve as a tibial bearing surface for articulation with the femoral component

Publicado el 27 de agosto de 2014 · Inicio del retiro: 11 de julio de 2014

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Ortho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK Tibil Insert because the foil pouches did not properly seal, allowing air to seep back into the pouch after sealing which increases the risk that the material properties of the implant would be substantially degraded due to oxidation.

Qué hacer

Voluntary: Firm initiated · E-Mail

Detalles del retiro

Fabricante
Ortho Development Corporation
Distribuidor
Worldwide Distribution - US Nationwide and the country of Japan. No military/govt/VA consignees.
Lote
UC Tibial Insert Sz 3 10mm,Part 163-3310,Lot A136109,UC Tibial Insert Sz 4 8mm,Part 163-3408,Lot A135948,UC Tibial Insert Sz 4 14mm,Part 163-3414,Lot A136110,UC Tibial Insert Sz 6 7mm,Part 163-3607,Lot A136113,UC Tibial Insert Sz 6 7mm,Part 163-3607,Lot A136746,
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States