FDAFDA · Alimentos y medicamentos·N.º oficial Z-2437-2020
(1) CentraLink Data Management System software version 16.0.2 Siemens Material Number (SMN): 11313246 (2) CentraLink Data Management System software version 16.0.3 Siemens Material Number (SMN): 11314337 Product Usage: are clinical laboratory information modules which consolidate test orders and test results, provide test ordering and priority routing information to automation, apply rules to test orders and auto-verification rules for test results, and provide integrated management of quality control of test system.
Publicado el 8 de julio de 2020 · Inicio del retiro: 15 de mayo de 2020
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Unexpected interface driver behavior identified (QC) Results May Be Assigned to an Incorrect Control Lot Number, and lead to the reporting of erroneous patient results if the QC failed but appeared to be passing and the issue is not detected during QC review
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Siemens Healthcare Diagnostics, Inc.
- Distribuidor
- Worldwide distribution - US Nationwide distribution and the countries of Argentina, Australia, Bahamas, Belgium, Brazil, Chile, Colombia, Czech Republic, Egypt, Germany, Greece, Hungary, India, Italy, Kuwait, Latvia, Lithuania, Malaysia, Mexico, Oman, Peru, Poland, Portugal, Qatar, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Thailand, Turkey, United Arab Emirates, United Kingdom.
- Lote
- Software versions: CentraLink Data Management System software version 16.0.2 UDI: 00630414220635 CentraLink Data Management System software version 16.0.3 UDI: 00630414237268
- Clasificación de riesgo
- Class II
- Área (FDA)
- device