SPECULAR MICROSCOPE CEM-530; Software version 1.08 and 1.09. Opthalmic: The NIDEK Specular Microscope CEM-530 provides non-contact, high magnification image capture of endothelium enabling observation of the size and shape of cells. Information such as the number of endothelial cells, cell density, and cell area is analyzed through the captured images. The captured images and analysis results of endothelium are used for intraocular or corneal surgery, postoperative follow-up and corneal observation such as for endothelial disorders or the corneal state of patients who wear extended-wear contact lenses.

Publicado el 7 de septiembre de 2016 · Inicio del retiro: 8 de agosto de 2016

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Software version 1.08 and 1.09 for the Specular Microscope CEM 530 included a change of analysis results feature that was not reviewed and approved by the FDA.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Nidek Inc
Distribuidor
US Distribution to states of: TX, KS, FL, NV, CA, PA, KY, OR, TN, AL, VA, MO, OK, CT, NY. LA, NM, and IN.
Lote
Microscopes with serial numbers: 120004,120005, 120006, 120007, 120009, 120010, 120011,120012, 120013, 120014, 0120016, 120019, 120021, 120023, 120024, 120026, 120030, 120036, 0120038, 120039, 120040, 120042, 120044,120045, 120046, 120047, 120048, 120049, 120050.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States