Cobas 8000 Modular Analyzer Series, Model/Catalog/Part Number: 05641489001. The Cobas 8000 is a fully automated system for clinical chemistry analysis.
Publicado el 12 de diciembre de 2012 · Inicio del retiro: 1 de noviembre de 2012
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
On cobas c 701 and cobas c 702 modules of the Cobas 8000 series analyzer, a mismatch between the reagent probe positioning and the aspirated reagent volumes under certain conditions can lead to insufficient reagent pipetting. This may affect patient sample and controls recovery for the following tests: Alkaline Phosphatase (ALP2), Alanine Aminotransferase (ALT), Bicarbonate (CO2-L), Creatinine (C
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Roche Diagnostics Operations, Inc.
- Distribuidor
- Nationwide Distribution, including the states of AL, AZ, CA, FL, GA, IL, IN, KS, KY, MA, MD, MI, MN, NC, NJ, NY, OH, VA, WA, and WI.
- Lote
- Catalog numbers affected: Alkaline Phosphatase acc. to IFCC Gen.2 05166888190; Alanine Aminotransferase acc. to IFCC without pyridoxal phosphate (P5P) activation 05850797190; Bicarbonate Liquid 05446376190; Creatinine Jaff¿ Gen.2 05168597190; C-Reactive Protein Gen.3 05172373190; and Urea/BUN 05171873190. Serial Number(s): 0917-10, 1020-04, 1020-05, 1021-03, 1021-04, 1021-07, 1021-08, 1023-06, 1023-09, 1023-10, 1024-03, 1024-04, 1024-05, 1024-09, 1025-02, 1025-05, 1025-07, 1025-09, 1026-01, 1026-02, 1027-09, 1028-01, 1028-07, 1030-09, 1031-05, 1031-09, 1032-09, 1032-10, 1033-04, 1033-05, 1107-06, 1109-02, 1109-03, 1111-07, 1111-08, 1112-02, 1112-04, 1115-06, 1115-07, 1115-08, 1115-10, 1117-04, 1117-07, 1118-02, 1126-02, 1126-03, 1126-04, 1127-01, 1127-02, 1127-06, 1127-07, 1138-03, 1138-04, 1139-04, 1139-05, 1139-06, 1140-02, 1140-03, 1140-04, 1140-05, 1140-10, 1141-08, 1141-09, 1142-02, 1142-03, 1142-04, 1143-01, 1143-02, 1143-03, 1143-04, 1143-05, 1143-06, 1144-05, 1146-04, 1146-05, 1146-06, 1146-07, and 1146-08.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States